Catalog No: RE1312M
Mouse Bcl-10(B-cell Lymphoma/Leukemia 10) ELIS Kit ELISA Kit
Copy Product Information

48T

$299.00

96T

$399.00

Product Details

Species:

Mouse

Alternative Name:

B-cell CLL/lymphoma 10CIPERCLAPB-cell lymphoma/leukemia 10, Bcl10, Bcl-10, CARD-containing apoptotic signaling protein, CARD-containing molecule enhancing NF-kappa-B, CARD-containing proapoptotic protein, CARD-like apoptotic protein, CARMEN, caspase-recruiting domain-containing protein, cCARMEN, c-E10, c-E10bcl-10, CED-3/ICH-1 prodomain homologous E10-like regulator, Cellular homolog of vCARMEN, cellular-E10, CIPER, CLAP, hCLAP, Mammalian CARD-containing adapter molecule E10, mE10, mE10CARD containing molecule enhancing NF-kB

Uniport ID:

Q9Z0H7

Assay Type:

Sandwich

unit:

ng/mL

Standard:

100ng/mL

Sensitivity:

0.94 ng/mL

Detection Range:

1.57-100ng/mL

Assay Time:

3.5H

Sample Type:

Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=100μL

Research:

Developmental Biology

Test Principle:

This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Mouse Bcl-10. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse Bcl-10 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse Bcl-10, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Mouse Bcl-10.You can calculate the concentration of Mouse Bcl-10 in the samples by comparing the OD of the samples to the standard curve.

Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
ng/mL OD Corrected
100 2.346 2.311
50 1.544 1.509
25 0.941 0.906
12.5 0.509 0.474
6.25 0.254 0.219
3.13 0.139 0.104
1.57 0.078 0.043
0 0.035 0

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 2.47 24.67 46.51 2.81 14.83 40.36
Standard deviation 0.09 0.83 2.38 0.12 1.22 2.64
C V (%) 7.11 3.48 7.2 3.73 6 4.25

Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type Range(%) Average Recovery(%)
Serum (n=8) 85-97 91
EDTA plasma (n=8) 80-94 87
Cell culture media (n=8) 91-105 98

Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 85-101 88-97 89-97
Average (%) 99 92 90
1:4 Range (%) 80-96 88-97 88-104
Average (%) 92 88 95
1:8 Range (%) 80-98 91-103 87-96
Average (%) 89 97 92
1:16 Range (%) 93-102 86-92 85-93
Average (%) 100 92 92

Assay Procedures

Description